Measurement | |||
Dew point | |||
Accuracy | ± 1 °C Td (0 ... 20 °C Td) ± 2 °C Td (-70 ... 0 / +20 … +50 °C Td) ± 3 °C Td (-100 ... -70 °C Td) | ||
Selectable units | %rH, °C Td, g/m³, mg/m³, g/m³ atm. ,mg/m³ atm. , ppmv, g/kg, °C Td atm. | ||
Measuring range | Sensor A: -100 ... +20 °C Td Sensor B: -50 ... +50 °C Td | ||
Repeatability | 0.5 °C | ||
Sensor | Sensor A: QCM + Polymer Sensor B: Polymer | ||
Pressure | |||
Accuracy | 0.5 % FS | ||
Measuring range | 0 ... 1.5 MPa (g) | ||
Sensor | Piezo resistive sensor | ||
Temperature | |||
Accuracy | ± 0.3 °C | ||
Measuring range | -30 ... +50 °C | ||
Sensor | PT 100 |
Interface & Supply | |||
Supply | |||
Power supply | USB charger: 5 V, 3 A Connector: USB-C | ||
Operating time | 8h. | ||
Data interface | |||
Connection | USB |
General data | ||
Display | ||
Integrated | 3.5” color LCD touch screen | |
Data Logger | ||
Storage | Integrated mass storage, 100 Million values | |
Material | ||
Housing | PC + ABS | |
Metal parts | Aluminium | |
Miscellaneous | ||
Protection class | IP30 | |
Approvals | CE | |
Weight | 2.7 kg complete set in transport case | |
Operating conditions | ||
Medium | Air, N₂, O₂, Argon, CO2 Note: The CO2 measurements with the A1371 sensor are limited to -40 °C Td. | |
Medium temperature | -30 ... +50 °C | |
Medium humidity | 0 ... 90 %, no condensation | |
Operating pressure | -0.1 ... 1.6 MPa (g)* | |
Ambient temperature | 0 ... +40 °C | |
Ambient humidity | 0 ... 80 % rH | |
Storage temperature | -20 ... +50 °C | |
Transport temperature | -30 ... 70 °C |
* At least 0.3 MPa(g) is needed for the measuring chamber supplied with the instrument.