Measurement | ||
Dew Point | ||
Accuracy | ± 1 ºC Td (0 ... 20 ºC Td) ± 2 ºC Td (-60 ... 0 ºC Td) ± 3 ºC (-100 ... -60 ºC Td) | |
Selectable units | °C, °F, bar(g), MPa(g), psi(g), % rH, g/m³, mg/m³, G/m³ atm., mg/m³ atm., ppm(v), g/kg, °C Td atm. | |
Measuring range (model depending) | S230: -100 ... +20 °C Td S231: -50 ... +20 °C Td | |
Sensor | Polymer + QCM | |
Response time t90 (@ 4l/min) | 20 °C Td -> -60 °C Td = < 240 sec -60 °C Td -> -20 °C Td = < 30 sec | |
Temperature | ||
Accuracy | 0.3 °C | |
Measuring range | -30 ... +70 °C | |
Sensor | NTC | |
Reference conditions | ||
Selectable conditions | Pressure Dew Point, Atmospheric Dew Point |
Signal / Interface & Supply | ||
Analog output | ||
Signal | ± 1 ºC Td (0 ... 20 ºC Td) ± 2 ºC Td (-60 ... 0 ºC Td) ± 3 ºC (-100 ... -60 ºC Td) | |
Scaling | S230: 4 mA = -100 °C Td; 20 mA = +20 °C Td S231: 4 mA = -50 °C Td; 20 mA = +20 °C Td | |
Load | 250R | |
Update rate | 3/sec | |
Fieldbus | ||
Protocol | Modbus/RTU | |
Update rate | 1/sec | |
Supply | ||
Voltage supply | 15 ... 30 VDC | |
Current consumption | 40 mA @ 24 VDC |
General data | ||
Configuration | ||
PC Software | S4C-DP Application | |
Material | ||
Process connection | Stainless steel 1.4301 (SUS 304) | |
Housing | Aluminium alloy | |
Sensor | Stainless steel 1.4301 (SUS 304) | |
Metal parts | Sinter Filter (stainless steel) | |
Miscellaneous | ||
Electrical connection | Screw terminals | |
Protection class | IP65 | |
Approvals | Ex db[ib] IIC T4 Gb, CE | |
Process connection | G 1/2“ thread (ISO 228/1) | |
Weight | 728 g | |
Operating conditions | ||
Medium | Air, Argon, O2, N2, CO2* | |
Medium quality | ISO 8573-1: 4.6.3 or better | |
Medium temperature | 0 ... +50 °C | |
Medium humidity | ≤ 20 °C Td | |
Operating pressure (model depending) | -0.1 ... 1.6 MPa (S230) -0.1 ... 35 MPa (S231) | |
Ambient temperature | 0 ... +50 °C | |
Ambient humidity | 0 ... 100 % rH | |
Storage temperature | -20 ... +50 °C | |
Transport temperature | -30 ... +70 º |
Stated accuracy under following conditions:
• Ambient temperature 23 °C ±3 °C
• Process temperature 23 °C ±3 °C
• Ambient humidity < 95 %, no condensation
• Airflow > 2 l/min at sensor tip
문의 031)243-1035
FAX 031)243-1025
sales@gms-korea.com