Technical Data ‹red-y smart series› | ||||
Instrument types |
|
| ||
smart meter GSM Thermal mass flow meter | smart controller GSC Thermal mass flow controller | OEM version For customer-specific requirements | ||
Instrument versions | ||||
‹Standard› The economic solution | Accuracy: ± 1.0 % of full scale(1) Turndown ratio: 1 : 50 | |||
‹Hi-Performance› With highest accuracy and turndown ratio (available for GSM < 200 ln/min / GSC < 150 ln/min (air)) | Accuracy: ± 0.3 % of full scale + ± 0.5% of reading(1) Turndown ratio: 1 : 100 1) An additional error of ±0.25% may apply for analogue signals | |||
Measuring ranges | ||||
(Air/Full scale freely selectable) | Type | Measuring range (air) | Connection | |
red-y smart meter GSM Meter | GSM-A GSM-B GSM-C GSM-D | from 0 ... 25 mln/min to 0 ... 600 mln/min from 0 ... 600 mln/min to 0 ... 6000 mln/min from 0 ... 6 ln/min to 0 ... 60 ln/min from 0 ... 60 ln/min to 0 ... 450 ln/min
| G¼" G¼" G¼" G½" | |
red-y smart controller GSC Controller | GSC-A GSC-B GSC-C GSC-D | from 0 ... 25 mln/min to 0 ... 600 mln/min from 0 ... 600 mln/min to 0 ... 6000 mln/min from 0 ... 6 ln/min to 0 ... 60 ln/min from 0 ... 60 ln/min to 0 ... 450 ln/min | G¼" G¼" G¼" G½" | |
Performance data | ||||
Media (real gas calibration) | Air, O2(2), N2(2), He, Ar, CO2, H2, CH4, C3H8 (other gases and gas mixtures on request) 2)O2 & N2 are calibrated with air | |||
Response time | Meter (GSM): ± 80ms(3); Controller (GSC): ± 500ms(3) 3)depending on device configuration & according to SEMI standard E17-1011, 5-100% of range under optimized conditions | |||
Repeatability | ± 0.2% of full scale (according to SEMI standard E56-0309) | |||
Longterm stability | < 1% of measured value / year | |||
Power supply | 24 Vdc (18 – 30 Vdc), 15 Vdc on request | |||
Current consumption Standard | Meter (GSM): max. 100mA; Controller (GSC): max. 250mA (GSC with valve type 8 max. 490mA) | |||
Current consumption Profinet RT / EtherCAT | Meter (GSM): max. 125mA; Controller (GSC): max. 340mA (GSC with valve type 8 max. 560mA) | |||
Operation pressure | 0.2 – 11 bar a (GSC with valve type 4.5 and 8 max. 8 bar a) | |||
Temperature (environment/gas) | 0 – 50°C | |||
Materials | Anodized aluminium, optional stainless steel electropolished | |||
Seals | FKM, EPDM, optional FFKM | |||
Pressure sensitivity | < 0.2% / bar of reading (typical N2) | |||
Temperature sensitivity | < 0.025% FS measuring range type / °C | |||
Warm-up time | < 1 sec. for full accuracy | |||
Integration | ||||
In- / Output signals analog | 0..20 mA, 4..20 mA, 0..5 V, 1..5 V, 0..10 V, 2..10 V | |||
In- / Output signals digital | RS-485; Modbus RTU (Slave); Lab View-VIs available Option: ProfiBus DP-V0, DP-V1 / Profinet RT / EtherCAT | |||
Process connection | G¼" (BSPP(4) female) up to 60 ln/min, G½" (BSPP(4) female) up to 450 ln/min 4)British Standard Pipe Paralle | |||
Inlet section | None required | |||
Electrical connection | Sub D plug, 9 pole Option ProfiBus: Sub D 9 pole / Option Profinet RT or EtherCAT: 2x RJ45 (IN/OUT) | |||
Mounting orientation | Any position (consult manufacturer above 5 bar or vertical mounting) | |||
Safety | ||||
Test pressure | 16 bar a | |||
Leak rate | < 1 x 10-6 mbar l/s He | |||
Ingress protection class | IP-50 | |||
EMC | EN 61326-1 |
문의 031)243-1035
FAX 031)243-1025
sales@gms-korea.com